PRODUCT INFORMATION

e-Myco™ VALiD-Q Real-time PCR Kit (v3.0 PRO)

ManufactureriNtRON Biotechnology Catalogue / SKU25247
Price £1,204.95 GBP UK pricing

UK/EU ordering: add this item to cart, review your order, then continue securely to checkout.

PRODUCT INFORMATION

Clear technical information,
without the clutter.

Product typePCR & qPCR Reagents CollectionPCR & qPCR Reagents
01 Description, applications & specifications

Description

The e-Myco™ VALiD-Q Real-time PCR Kit (v3.0 PRO) provides a rapid, reliable, and guideline-compliant method to detect mycoplasma contamination in cell-based manufacturing processes. The assay offers comprehensive coverage of international pharmacopeia standards (EP, USP, JP, and Korea MFDS), enabling simultaneous detection of compendial reference species within a single integrated panel. Panelized primers and probes, developed through extensive sequence alignment and in-silico validation, ensure high specificity and meet established detection limits without cross-reactivity with non-target microorganisms.


▶ Full compliance with Major 4 pharmacopeias : A performance class beyond comparison

▶ Just 53 minutes : Far beyond the limitations of culture methods

▶ Internal Control (IC) and External Control (EC) ensure process validity across all steps, from nucleic acid extraction to PCR amplification

Major Mycoplasma Species Detectable According to Guidelines
A, (Alternative reference species) : In the European Pharmacopoeia (EP) and Korea MFDS, it is available as Mycoplasma pneumoniae or Mycoplasma gallisepticum.

 B, (Insect, Plant, or Avian-derived species) : In the Japanese Pharmacopoeia (JP), species of insect or plant origin (e.g., Spiroplasma citri) should also be tested in addition to the listed species. When avian cells or materials are used, avian-derived species (e.g., Mycoplasma synoviae) may also be considered for inclusion in the test panel.

​


Evaluation of Detection Performance at 10 CFU/mL under Rapid Testing Conditions. The e-Myco™ VALiD-Q Real-Time PCR Kit (v3.0 PRO), which can perform mycoplasma testing in less than 1 hour, is faster than competitive products and has been proven to achieve 100% detection using a certified mycoplasma reference standard panel with a concentration of 10 CFU/mL.​

Key Features

▶ Validation for the production of Biopharmaceuticals & Advanced Therapy Medicinal Products(ATMP)

▶ Fully compliant with EP 2.6.7, USP <63>, JP G3, Korea MFDS Mycoplasma NAT guidelines

▶ 53 minute real-time PCR detection of Mycoplasma and related Mollicutes class

▶ Targets a highly conserved 16S rRNA gene region for sensitive qualitative detection

▶ Application of IC & EC system to verify the validity of qPCR

Storage

▶ Storage Temperature : Below - 20°C

▶ Shelf Life : 1 year (Stable Within 6 months after opening, within expiry date of the kit) ​

Kit Contents

No. Kit Contents Unit
1 2X qPCR mix 500 μl × 2 tubes
2 Detection solution 280 μl × 1 tube
3 Internal Control 150 μl × 2 tubes
4 External Control 25 μl × 3 tubes
5 DNase/RNase free water 1,000 μl × 1 tube
6 Quick guide 1 ea

Technical Data

1. Limit of Detection (LOD) test



● Detection limit evaluation of the e-Myco™ VALiD-Q Real-time PCR Kit (v3.0 PRO)​To evaluate the detection sensitivity of the assay, genomic DNA from ten quantified Mycoplasma strains(A. laidlawii, M. arginini, M. gallisepticum, M. fermentans, M. hyorhinis, M. orale, M. pneumoniae, M. salivarium, M. synoviae, and S. citri) was serially diluted to 10 CFU/mL and 1 CFU/mL. Nucleic acids were extracted directly from culture media and tested using the e-Myco™ VALiD-Q Real-time PCR Kit (v3.0 PRO). All species showed positive detection at the lowest dilution level, confirming a detection limit of 1 CFU/mL for each Mycoplasma species tested.

2. Specificity test


● ​Analytical specificity evaluation of the e-Myco™ VALiD-Q Real-time PCR Kit (v3.0 PRO)​ The analytical specificity was evaluated against six non-Mycoplasma strains according to the European Pharmacopoeia (EP) 2.6.7 guidelines. No amplification signals were observed in the non-target strains, confirming high analytical specificity.

3. Robustness test


● ​Reliability and reproducibility of the Mycoplasma Target (FAM) assay under different instrument, sample matrix,extraction methods, PCR reagents conditions​ Performance was evaluated for A. laidlawii (10 CFU/mL) across four experimental variables PCR instruments, nucleic acid extraction methods, sample matrices, and PCR conditions. Under all conditions, Ct values remained within the expected range, and the coefficient of variation (CV) was ≤5%, demonstrating stable performance. 

4. Dual-Fluorescence Control Ensuring Confidence in NAT Results


● ​Dual-channel External Control (EC) system for identifying true positives and carry-over contamination in the e-Myco™​ VALiD-Q Real-time PCR Kit (v3.0 PRO)​ Amplification curves represent : (1) True Positive (Sample), FAM-only amplification, (2) External Control Positive, FAM + Cy5 amplification generated from the EC template, (3) False Positive (PC carry-over), dual FAM/Cy5 fluorescence detected in a test sample due to contamination. Dual-channel positivity in a sample indicates carry-over of EC or positive-control material; such reactions are invalid and must be repeated.​

Available pack options: 100 Tests. Manufacturer reference: 25247.